Abstract:
Episodes of patient-ventilator asynchrony (PVA) occur during acute and chronic non-invasive positive pressure ventilation (NIV). In long-term NIV, description and quantification of PVA is not standardised, thus limiting assessment of its clinical impact. The present report provides a framework for a systematic analysis of polygraphic recordings of patients under NIV for the detection and classification of PVA validated by bench testing. The algorithm described uses two different time windows: rate asynchrony and intracycle asynchrony. This approach should facilitate further studies on prevalence and clinical impact of PVA in long-term NIV.
Tipologia CRIS:
Articolo su Rivista
Keywords:
asynchrony; central sleep apnea; home ventilators; monitoring; non-invasive ventilation; obstructive sleep apnoea; polysomnography; sleep-disordered breathing
Elenco autori:
Gonzalez-Bermejo, J.; Janssens, J. -P.; Rabec, C.; Perrin, C.; Lofaso, F.; Langevin, B.; Carlucci, A.; Lujan, M.