Lattice parameters determination from powder diffraction data: Results from a round robin project
Articolo
Data di Pubblicazione:
1996
Abstract:
A well-crystallized sample of powdered KCl has been distributed among several laboratories in order to test the reproducibility of the lattice parameter measurement on different X-ray powder diffraction instruments. The precision of the determined unit-cell dimension is in the 10-5 Å range, while the discrepancies among the results from different laboratories using the same numerical analysis are at least one order of magnitude higher. It is shown that if different numerical analyses, including full pattern refinement, are used, values differing in the third decimal digit are obtained for the same data set. © 1996 International Centre for Diffraction Data.
Tipologia CRIS:
Articolo su Rivista
Keywords:
Accuracy; Lattice parameters; Potassium chloride; Precision; Round robin
Elenco autori:
Masciocchi, N.; Artioli, G.
Link alla scheda completa:
Pubblicato in: