Data di Pubblicazione:
1997
Abstract:
A silicon strip detector for the SYRMEP (SYncrotron Radiation for MEdical Physics) experiment has been designed and realised. The main features of this detector are AC-coupling through integrated coupling capacitors, DC biasing by means of a gated punch-through structure, bulk contact on the junction side through a forward-biased p(+) implant and integrated fan-in on active silicon. Results of laboratory tests of the detector parameters, allowing a thorough evaluation of the technological solutions employed, are presented.
The functionality of the detectors and the charge collection linearity have been tested with different gamma sources, using hybrid, low-noise front-end electronics.
The functionality of the detectors and the charge collection linearity have been tested with different gamma sources, using hybrid, low-noise front-end electronics.
Tipologia CRIS:
Articolo su Rivista
Keywords:
SYRMEP; digital mammography; silicon detector; readout ASIC
Elenco autori:
Arfelli, F; Barbiellini, G; Bonvicini, V; Bravin, A; Cantatore, G; Castelli, E; Dimichiel, M; Longo, R; Olivo, A; Pani, S; Pontoni, D; Poropat, P; Prest, Michela; Rashevsky, A; Tromba, G; Vacchi, A; Zampa, N.