Design and evaluation of AC-coupled, FOXFET-biased, ''edge-on'' silicon strip detectors for X-ray imaging
Articolo
Data di Pubblicazione:
1997
Abstract:
A silicon strip detector for the SYRMEP (SYnchrotron Radiation for MEdical Physics) experiment has been designed and realised. The main features of this detector are AC-coupling through integrated coupling capacitors, DC bias of the strips by means of a gated punch-through structure, bulk contact on the junction side through a forward-biased p(+) implant, thinned entrance window for the incoming radiation (in an ''edge-on'' geometry) and integrated fan-in on active silicon. Results of laboratory tests of the detector parameters, allowing a thorough evaluation of the technological solutions employed, are presented.
Tipologia CRIS:
Articolo su Rivista
Keywords:
SYRMEP; digital mammography; silicon strip detector; AC coupled; FOXFET biased
Elenco autori:
Arfelli, F; Bonvicini, V; Bravin, A; Burger, P; Cantatore, G; Castelli, E; Dimichiel, M; Longo, R; Olivo, A; Pani, S; Pontoni, D; Poropat, P; Prest, Michela; Rashevsky, A; Tromba, G; Vacchi, A; Zampa, N.