Mammography of a phantom and breast tissue with synchrotron radiation and a linear-array silicon detector
Articolo
Data di Pubblicazione:
1998
Abstract:
A linear-array, silicon pixel detector, capable of counting single photons, was applied to mammography by using a synchrotron radiation beam. Images were obtained of both a mammographic phantom and a breast-tissue sample. The phantom image was acquired with a mean glandular dose of 0.32 mGy. This detector combined with a synchrotron radiation beam allows acquisition of high-contrast, low-dose images of soft tissues.
Tipologia CRIS:
Articolo su Rivista
Keywords:
Digital mammography; silicon pixel detector; single photon counting
Elenco autori:
Arfelli, F; Bonvicini, V; Bravin, A; Cantatore, G; Castelli, E; Dalla Palma, L; Di Michiel, M; Longo, R; Olivo, A; Pani, S; Pontoni, D; Poropat, P; Prest, Michela; Rashevsky, A; Tromba, G; Vacchi, A.